[1]廖文胜)△,刘玉峰),鲍恒),等.经口前路寰枢椎侧块关节间固定融合器的研制及生物力学分析[J].郑州大学学报(医学版),2013,(05):658.
 LIAO Wensheng),LIU Yufeng),BAO Heng),et al.Biomechanical analysis of anterior transoral fixation and fusion device for atlantoaxial lateral mass joints[J].JOURNAL OF ZHENGZHOU UNIVERSITY(MEDICAL SCIENCES),2013,(05):658.
点击复制

经口前路寰枢椎侧块关节间固定融合器的研制及生物力学分析
分享到:

《郑州大学学报(医学版)》[ISSN:1671-6825/CN:41-1340/R]

卷:
期数:
2013年05期
页码:
658
栏目:
应用研究
出版日期:
2013-09-02

文章信息/Info

Title:
Biomechanical analysis of anterior transoral fixation and fusion device for atlantoaxial lateral mass joints
作者:
廖文胜1)刘玉峰2)鲍恒1)王利民1)
1)郑州大学第一附属医院骨科 郑州 4500522)洛阳正骨医院骨质疏松科 洛阳471002
Author(s):
LIAO Wensheng1) LIU Yufeng2) BAO Heng1) WANG Limin1)
1)Department of Orthopedics, the First Affiliated Hospital, Zhengzhou University, Zhengzhou 450052 2)Department of Osteoporosis, Luoyang OrthopedicTraumatological Hospital,Luoyang 471002
关键词:
颅底凹陷症寰枕融合经口咽入路关节融合术内固定器
Keywords:
basilar invagination occipital atlas fusion transoral pharyngeal approach arthrodesis internal fixator
分类号:
R687.4
摘要:
目的:研制经口前路寰枢椎侧块关节间固定融合器,并进行生物力学评价。方法:基于解剖学观察,设计和制作该融合器。取7例成年颈椎标本,先采用固定融合器固定,然后用万能生物力学试验机进行生物力学分析,记录30、50和70 N载荷下的轴向位移。拆卸器械,用后路寰枢椎钉棒固定后,进行生物力学分析。结果:30、50和70 N载荷下,前路固定融合器固定后标本轴向位移分别为(0.39±0.07)、(0.76±0.05)、(1.15±0.09) mm,后路钉棒固定后的轴向位移分别为(1.06±0.09)、(1.53±0.11)和(1.99±0.10) mm。相同载荷下,前路固定融合器固定的轴向位移均小于后路钉棒固定方式(t=11.130,17.120和16.150,P<0.001)。结论:设计的经口前路寰枢椎侧块关节间固定融合器具有更好的轴向稳定性。
Abstract:
Aim: To design a new anterior transoral fixation and fusion device for arthrodesis of atlantoaxial joints and test it biomechanically. Methods:Based on the previous anatomical study of the human cervical specimens, a new device for anterior transoral fixation and fusion of atlantoaxial lateral mass joints was developed. Seven human cervical specimens (C1C2) were fixed with this device,and the axial displacement under the load of 30 N, 50 N, and 70 N were tested biomechanically with a universal testing machine,then the specimens were removed the device and fixed again with the posterior screwrod system including C lateral mass screws, C pedicle screws and rods,and were tested again. Results: Under the load of 30, 50,and 70 N,the axial displacements were (0.39±0.07),(0.76±0.05), and (1.15±0.09) mm when fixed with the new device,and (1.06±0.09),(1.53±0.11), and (1.99±0.10) mm when fixed with the posterior screwrod system.Under the same load,the axial displacements of the new device fixation were less than those of the posterior screwrod system(t=11.130,17.120 and 16.150,P<0.001). Conclusion: Anterior transoral fixation and fusion device has higher axial stability.

参考文献/References:

[1]Hosalkar HS,Sankar WN,Wills BP,et al.Congenital osseous anomalies of the upper cervical spine[J].J Bone Joint Surg Am,2008,90(2):337
[2]王超,阎明,周海涛,等.难复性寰枢关节脱位的手术治疗[J].中华骨科杂志,2004,24(5):290
[3]Deutsch H,Haid RW Jr,Rodts GE Jr,et al.Occipitocervical fixation: longterm results[J].Spine(Phila Pa 1976),2005,30(5):530
[4] 闫明,王超,王圣林,等.经口咽入路寰枢关节松解复位术后发生的败血症(附4 例报道)[J].中国脊柱脊髓杂志,2009,9(12):920
[5]Schmelzle R, Harms J, Stoltze D. Osteosynthesen im occipitocervicalemu bergang vom transoralen Zugang aus[M]//XVII SICOT World Congress Abstracts. Munich: DemeterVerlag,1987:27
[6]Kandziora F, Pflugmacher R, Ludwig K, et al. Biomechanical comparison of four anterior atlantoaxial plate systems[J]. J Neurosurg,2002,96(Suppl 3):313
[7] Yin Q,Ai F,Zhang K,et al.Irreducible anterior atlantoaxial dislocation: onestage treatment with a transoral atlantoaxial reduction plate fixation and fusion.Rreport of 5 cases and review of the literature[J].Spine (Phila Pa 1976),2005,30(13):E375
[8]Ai F,Yin Q,Wang Z,et al.Applied anatomy of transoral atlantoaxial reduction plate internal fixation[J].Spine(Phila Pa 1976),2006,31(2): 128
[9]Li S,Ni B, Xie N, et al.Biomechanical evaluation of an atlantoaxial lateral mass fusion cage with C1C2 pedicle fixation[J].Spine (Phila Pa 1976),2010,35(14):E624
[10]胡勇,谢辉,杨述华.人工寰齿关节生物力学及临床前期研究进展[J].医用生物力学,2006,21(4):334
[11]Wilke HJ, Wenger K, Claes L.Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants[J].Eur Spine J,1998,7(2):148